Verification Process - Invacare Homefill Cylinder - Invacare HomeFillII AW Service Manual

Table of Contents

Advertisement

SECTION 14—MAINTENANCE

Verification Process - Invacare HomeFill Cylinder

NOTE: Inspection to be performed upon delivery of unit from Invacare and upon return of unit
from rental periods.
MODEL No.________________________
ON EACH INSPECTION
Record Date of Service
Clean and disinfect the equip-
ment per Cleaning/Disinfection
policy
CAUTION
Hand wipe conserver
and regulator. DO NOT
soak or spray fluids
directly onto conserver
or regulator.
Visual Inspection
Check the cylinder Hydrostatic
test date. If the cylinder is within
6 months of needing to be re-
certified, remove the cylinder
from service and send it to be
Hydrostatic tested/re-certified
Bleed the cylinder down until it is
empty (use 2 L/min. continuous
flow setting on conserver)
Examine the cylinder for: Dents
or Dings, Arc Burns, Oil or
Grease, Any other signs of dam-
age, Any signs of Fire or Heat
damage, Presence of legible label
Inspect the valve for: Debris, Oil,
or Grease, Noticeable signs of
damage, Signs of corrosion inside
of the valve, Signs of heat damage
Inspect the gauge for visual dam-
age or needle inaccuracy. If dam-
aged or inaccurate, remove the
cylinder from service.
NOTE: If damage is found, return unit to Invacare.
Function Test (Procedure
below)
®
HomeFill
II AW Compressor
SERIAL No._______________________
98