Cylinder Fill Times; Leak Test - Invacare HomeFillII AW Service Manual

Table of Contents

Advertisement

Cylinder Fill Times

CYLINDER
FILLING TIMES
NORMAL
MAXIMUM
NOTE: Filling time is approximate and may vary with environmental conditions.
NOTE: *Not for EU countries

Leak Test

Use extreme caution when testing this device. It may start unexpectedly, and it
contains many pinch points.
DO NOT spray electrical connections with leak-tek.
NEVER allow HomeFill to continue running above 2100 psi (14479 kPa).
DO NOT tighten fittings beyond a ¼ turn. Tightening beyond ¼ turn risks
stretching and weakening the copper material and causing a ruptured fitting.
Tools and Supplies Needed:
• High Pressure Gauge
• Leak Tek 50/50 Mix
• Flashlight
1. Plug HomeFill into an electrical outlet.
2. Attach a high pressure gauge to the fill line disconnect.
3. Attach the oxygen concentrator line to the inlet fitting. Refer to Concentrator
Verification and Maintenance on page 16.
4. Ensure that the concentrator output is greater than 90% at 2.5 L/min for Platinum 5,
Platinum S or Perfecto
to Concentrator Verification and Maintenance on page 16.
5. Turn the power switch On.
6. Note that four of the five LEDs and the FILLING LED light at start-up.
7. If unit does not start within three minutes, confirm power and oxygen inputs are
present.
8. Once HomeFill starts, remove the input tubing. Unit should shut down within 30
seconds.
9. Reinstall the input tubing, unit should restart within 30 seconds.
Part No 1154316
SECTION 13—TROUBLESHOOTING/CYLINDER FILL TIMES
Concentrator Flow Rate to Patient:
Platinum 5, Platinum S, Perfecto
Platinum 9 up to 5 L/min
*HF2PC9
*HF2PCL4
2 hr 20 min
2 hr 45 min
WARNING
concentrators or 5 L/min for Platinum 9 concentrators. Refer
2
up to 2.5 L/min
2
*HF2PCL6
60 min
1 hr 25 min
1 hr 20 min
1 hr 50 min
87
HF2PCE8
2 hr 5 min
N/A
®
HomeFill
II AW Compressor