SECTION 8—BASE REPLACEMENT
SECTION 8—BASE REPLACEMENT
7
Tools Required:
/
open ended wrench,
16
Needle Nose Pliers, Phillips screwdriver.

Disassembly

NOTE: For this procedure, refer to FIGURE 8.1 through FIGURE 8.7.
1. Install an empty cylinder onto the compressor to release pressure in lines. Refer to
Connecting/Disconnecting Cylinder to/from Compressor on page 20.
2. Remove cylinder. Refer to Connecting/Disconnecting Cylinder to/from Compressor
on page 20.
3. Disconnect power from the HomeFill unit.
4. Remove the cabinet. Refer to Removing/Installing the Cabinet on page 25.
5. Disconnect the two pin connector that connects the motor to the wiring harness
(FIGURE 8.1).
6. Using needle nose pliers, disconnect the two wires from the capacitor (FIGURE 8.1).
7. Using needle nose pliers, disconnect the grounding clip from the motor (FIGURE 8.1).
Control Panel Gusset
Two Pin
Connector
(STEP 4)
Motor
Grounding
Clip
Wires (STEP 5)
®
HomeFill
II AW Compressor
5
Capacitor
Mounting Screw
(Hidden behind wiring assembly)
FIGURE 8.1 Base Replacement
/
open ended wrench, Diagonal Cutters,
8
44
Control Panel
Gusset
Mounting
Screw
Part No 1154316