SECTION 6—PANEL ASSEMBLY

Replacing/Rebuilding Compressor Assembly

There are no serviceable parts in the compressor assembly. The HomeFill II MUST be
returned to Invacare for service.

Replacing Accumulator Assembly

DANGER
To prevent electrical shock, ALWAYS disconnect compressor from electrical
outlet before servicing.
Tools required for this procedure are:
• Phillips Screwdriver

Units manufactured before 10/03/05

NOTE: For this procedure, refer to FIGURE 6.5 on page 30.
Removing Accumulator Assembly
1. Remove the cabinet. Refer to Removing/Installing the Cabinet on page 21.
2. Remove the control panel. Refer to Removing/Installing Control Panel on page 25.
3. Remove the two mounting screws securing the accumulator assembly to the main
brackets.
4. Cut the two tie-wraps securing the two fittings to the oxygen hoses.
5. Remove the existing accumulator assembly.
Installing Accumulator Assembly
1. Install the accumulator assembly to the main brackets with two mounting screws.
2. Connect the two fittings to the oxygen hoses.
3. Secure the two fittings to the oxygen hoses with two tie-wraps.
4. Install the control panel. Refer to Removing/Installing Control Panel on page 25.
5. Install the cabinet. Refer to Removing/Installing the Cabinet on page 21.
29
Part No 1141487
HomeFill™II Compressor

Hide quick links:

Advertisement

Table of Contents
loading