SECTION 10—COMPRESSOR REPLACEMENT
23. Remove the six mounting screws that secure the compressor/motor assembly to the
base (FIGURE 10.11).
24. Remove the compressor/motor assembly from the base.
TOP VIEW WITH CONTROL PANEL ASSEMBLY REMOVED
Base
Mounting
Screws
FIGURE 10.11 Compressor Replacement - HomeFill Units Manufactured After 8/1/05
25. Carefully place the pump assembly upside down on the table.
26. Remove the timing belt by gently pulling the belt away from the large pulley and
motor pulley. Refer to Detail "A" in FIGURE 10.12.
27. Rotate pump assembly 180° so that the motor is facing up.
28. Remove the two mounting screws, two bushings, and two washers that secure the
motor to the pump assembly. Refer to Detail "B" and "C" in FIGURE 10.12.
29. Remove the motor from the pump assembly.
NOTE: Retain the two mounting screws, two bushings, and two washers for reassembly.
30. Remove the mounting screw and washer (Detail "A") that secure the large pulley to
the pump assembly crankshaft. Refer to FIGURE 10.12 on page 73.
®
HomeFill
II AW Compressor
Mounting Screws
72
Motor/Pump Assembly
Mounting
Screws