31. Remove the retaining clip and hour meter from the base (FIGURE 8.6).
5
32. Using the
/
open ended wrench remove the nut from the receptacle fitting
8
(FIGURE 8.6).
33. Remove the receptacle fitting from the base (FIGURE 8.6).
Mounting
Hour Meter
Retaining Clip
Receptacle
Fitting
34. Perform one of the following (FIGURE 8.7):
• For units manufactured before 10/2/2005:
i. Remove the mounting screw that secures the bracket to the base.
ii. Remove pin partially out.
iii. Lift connector fillport assembly up off of base.
• For units manufactured after 10/2/2005:
i. Remove the two mounting screws that secure the bracket to the base.
ii. Remove pin partially out.
iii. Lift connector fillport assembly up off of base.
Part No 1154316
Mounting
Screw
Screw
Transformer
Power
Inlet
FIGURE 8.6 Base Replacement
SECTION 8—BASE REPLACEMENT
Mounting Screw
49
Base
Tie Wrap
Capacitor
®
HomeFill
II AW Compressor

Hide quick links:

Advertisement

Table of Contents
loading