Invacare® HomeFill® II Oxygen System

5.6 Replacing the Fuse

26
1.
Unplug the power cord from the electrical outlet.
2.
Remove the power cord from the power inlet A on the
compressor base C.
Pull out the fuse drawer B at the bottom of the power inlet.
3.
4.
Using a fuse puller or a small flat blade screw driver, remove the
front fuse inside the fuse drawer and discard.
The fuse drawer contains space for two fuses. The
front location is for the operational fuse. The unit is
supplied with the front fuse only. The rear location is
for an optional spare fuse. When replacing the front
fuse, ensure that there is a spare fuse with the same
type and rating in the fuse drawer.
5.
Install replacement fuse in front fuse D slot.
Use a 5 mm x 20 mm, time lag, low breaking capacity,
250V, 2 Amp, IEC 60127 style only (Invacare part
number 1099820).
6.
Push fuse drawer back into the power inlet.
7.
Install power cord into the power inlet on the compressor base.
8.
Plug power cord into electrical outlet.
9.
Turn the power switch to the On ( ) position and check to see
that the unit powers up properly.
10. If the unit does not come on, turn the power switch off, unplug
the power cord, and call your dealer, provider or Invacare for
service.