Invacare HomeFillII AW Service Manual page 45

Table of Contents

Advertisement

8. Disconnect the two wires from the high pressure switch (FIGURE 8.2).
7
9. Using the
/
open ended wrench, loosen the nut and disconnect the high pressure
16
oxygen line from the cylinder fillport connector (FIGURE 8.3).
10. Loosen the two screws that secure the bottom of the control panel in place
(FIGURE 8.3).
11. Remove the two screws that secure the control panel gussets to the rear of the base
assembly (FIGURE 8.1).
Mounting Screws
Part No 1154316
Wires
High Pressure Switch
FIGURE 8.2 Base Replacement
High Pressure Oxygen Line
FIGURE 8.3 Base Replacement
45
SECTION 8—BASE REPLACEMENT
Cylinder Fillport
Connection
Nut (Underneath)
®
HomeFill
II AW Compressor