Invacare HomeFillII AW Service Manual page 69

Table of Contents

Advertisement

8. Disconnect the two wires from the high pressure switch (FIGURE 10.8).
FIGURE 10.8 Compressor Replacement - HomeFill Units Manufactured After 8/1/05
7
9. Using the
/
open ended wrench, loosen the nut and disconnect the high pressure
16
oxygen line from the cylinder fillport connector (FIGURE 10.9).
10. Loosen the two screws that secure the bottom of the control panel in place
(FIGURE 10.9).
11. Remove the two screws that secure the control panel gussets to the rear of the base
assembly (FIGURE 10.7).
Mounting Screws
FIGURE 10.9 Compressor Replacement - HomeFill Units Manufactured After 8/1/05
DO NOT cut the tubing or damage (score) the tubing when removing tie-wraps.
Otherwise, HomeFill unit will not fill cylinders correctly.
Part No 1154316
SECTION 10—COMPRESSOR REPLACEMENT
Wires
High Pressure Switch
High Pressure Oxygen Line
CAUTION
69
Cylinder Fillport
Connection
Nut (Underneath)
®
HomeFill
II AW Compressor