SECTION 8—BASE REPLACEMENT
DO NOT cut the tubing or damage (score) the tubing when removing tie-wraps.
Otherwise HomeFill unit will not fill cylinders correctly.
12. Using diagonal cutters, cut the tie wrap that secures the oxygen inlet hose to the inlet
of stage 1 (FIGURE 8.4).
13. Disconnect the oxygen inlet hose from the inlet of stage 1 (FIGURE 8.4).
14. Using diagonal cutters, cut the tie wrap that secures the oxygen inlet hose to the
receptacle fitting (FIGURE 8.4).
15. Remove the oxygen inlet hose from the receptacle fitting (FIGURE 8.4).
16. Using diagonal cutters, cut the tie wrap that secures the oxygen sensor tubing to the
oxygen inlet hose (FIGURE 8.4).
17. Remove the oxygen sensor tubing from the oxygen inlet hose (FIGURE 8.4).
18. Note where each of the four wires and two jumpers are connected to the transformer.
19. Using needle nose pliers, disconnect the connectors from the transformer. See Detail
"A" in FIGURE 8.4.
20. Using needle nose pliers, disconnect the two connectors from the power inlet
(FIGURE 8.4).
21. Using needle nose pliers, disconnect the two connectors (not shown) from the hour
meter (FIGURE 8.4).
22. Lift control panel assembly off of base and set aside (FIGURE 8.4).
®
HomeFill
II AW Compressor
CAUTION
46
Part No 1154316