SECTION 8—BASE REPLACEMENT
Mounting Screw location
(For units manufactured
before 10/2/2005
Pin

Reassembly

NOTE: For this procedure, refer to FIGURE 8.1 through FIGURE 8.7.
1. Perform one of the following (FIGURE 8.7):
• For units manufactured before 10/2/2005:
i. Position the connector fillport assembly as shown.
ii. Secure the bracket to the base with a mounting screw.
iii. Push pin in.
• For units manufactured after 10/2/2005:
i. Position the connector fillport assembly as shown.
ii. Secure the bracket to the base with a mounting screw.
iii. Push pin in.
2. Install the receptacle fitting into the base and secure with nut. Securely tighten.
(FIGURE 8.6).
3. Install hour meter into the base and secure with retaining clip. (FIGURE 8.6)
4. Install the power inlet into the base and secure with two mounting screws
(FIGURE 8.6).
5. Install the transformer into the base and secure with two mounting screws.
(FIGURE 8.6)
6. Install the tie wrap mount (not shown) and secure in place with one mounting screw.
(FIGURE 8.6)
7. Position capacitor as shown and secure to tie wrap mount with tie wrap (FIGURE 8.6).
8. Secure the pump/motor assembly to the base with six mounting screws. Torque
mounting screws to 25-30 in-lbs (2.82-3.4 Nm) (FIGURE 8.5).
9. Position control panel assembly on base (FIGURE 8.4).
10. Connect the two connectors (not shown) to the hour meter (FIGURE 8.4).
®
HomeFill
II AW Compressor
FIGURE 8.7 Base Replacement
50
Mounting Screw location
(For units manufactured
after 10/2/2005
Part No 1154316