11. Connect the two connectors to the power inlet (FIGURE 8.4).
12. Connect the four connectors to the transformer as noted in STEP 18 on page 46.
13. Connect the oxygen sensor tubing to the oxygen inlet hose and secure with tie wrap
(FIGURE 8.4)
14. Connect the oxygen inlet hose to the receptacle fitting (FIGURE 8.4).
15. Secure the oxygen inlet hose to the receptacle fitting with a tie wrap (FIGURE 8.4).
16. Connect the oxygen inlet hose to the inlet of stage 1 (FIGURE 8.4).
17. Secure the hose with a tie wrap using a tie wrap gun set at #8 (FIGURE 8.4).
18. Secure the control panel gussets to the rear of the base assembly with two mounting
screws (FIGURE 8.1).
19. Secure the bottom of the control panel to the pump assembly with the two mounting
screws (FIGURE 8.3).
20. Connect the high pressure oxygen line to the cylinder fillport connector. Securely
tighten using the
21. Connect the two wires to the high pressure switch (FIGURE 8.2).
22. Connect the grounding clip to the motor (FIGURE 8.1)
23. Connect the two wires to the capacitor (FIGURE 8.1)
24. Connect the two pin connector that connects the motor to the wiring harness
(FIGURE 8.1).
25. Install the cabinet. Refer to Removing/Installing the Cabinet on page 25.
Part No 1154316
7
/
open ended wrench (FIGURE 8.3).
16
SECTION 8—BASE REPLACEMENT
51
®
HomeFill
II AW Compressor