Advertisement

This manual MUST be given to the user of the product.
BEFORE using this product, read this manual and save for future reference.
Invacare® HomeFill® Oxygen
System
Model IOH200
en Compressor
User Manual

Advertisement

Table of Contents
loading

Summary of Contents for Invacare HomeFill Oxygen System

  • Page 1 Invacare® HomeFill® Oxygen System Model IOH200 en Compressor User Manual This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference.
  • Page 2 All rights reserved. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by ™and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.
  • Page 3: Table Of Contents

    Contents 7 Troubleshooting ........27 7.1 Indicator Lights ....... . 27 7.2 Troubleshooting .
  • Page 4: General

    Invacare® HomeFill® Oxygen System Symbols on the Product 1 General The following symbols are found on the compressor, 1.1 Symbols compressor packaging or the oxygen cylinder. Symbols in the Manual Read the Manual (Compressor) Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage.
  • Page 5: Intended Use

    WARNING! Risk of Death, Injury or Damage Oxidizer The Invacare HomeFill compressor was designed to be used only with Invacare oxygen concentrators. Use with other types of devices Transport and Storage Temperature may cause injury, death or property damage. – DO NOT use the Invacare HomeFill compressor...
  • Page 6 Invacare® HomeFill® Oxygen System Platinum® — Top View Perfecto ™ — Rear View The intended use of an Invacare HomeFill compressor is an accessory device to an Invacare oxygen concentrator that enables the patient to refill custom portable oxygen cylinders for personal use.
  • Page 7: Safety

    – DO NOT use this product or any available – DO NOT use any lubricants on the compressor optional equipment without first completely unless recommended by Invacare. reading and understanding these instructions – NO SMOKING signs should be prominently and any additional instructional material such displayed.
  • Page 8 These Risk of Serious Injury or Damage substances may become flammable in the Use of non-Invacare accessories may result in presence of oxygen and cause injury. Avoid serious injury or damage. touching the cylinder fitting whenever possible.
  • Page 9 Safety WARNING! WARNING! Risk of Injury or Damage Risk of injury or damage A damaged product may malfunction and cause The HomeFill is equipped with a high pressure injury or damage. To avoid injury or damage: relief valve to ensure the user’s safety. When –...
  • Page 10: Label Locations

    Invacare® HomeFill® Oxygen System 2.2 Label Locations 1100873-M...
  • Page 11: Dealer Information

    Portable and mobile RF communications equipment can cord, otherwise damage may occur. affect the operation of this equipment. The compressor may be stacked on top on an Invacare HomeFill compatible In a double-insulated product, two systems of insulation oxygen concentrator during setup. The compressor operation are provided instead of grounding.
  • Page 12 • Reposition, relocate, or increase the separation between the devices. • Connect either line powered device to a different electrical power circuit. Contact Invacare for any additional EMC information or EMC tables for the device environment. 1100873-M...
  • Page 13: Setup

    3.2 Compressor Initial Setup Checklist Dealers ONLY — Refer to 2.3 Dealer Information, page 11 before performing this procedure. Users ONLY — Contact your Invacare dealer or Invacare for assistance with the setup procedure and to report unexpected operation or events.
  • Page 14: Choosing A Location

    Invacare® HomeFill® Oxygen System Connect the compressor to the concentrator. The compressor should be at least three inches away from walls, draperies or furniture to ensure sufficient air flow. 3.3 Choosing a Location Avoid placing the compressor next to heaters, radiators or hot air registers.
  • Page 15: Installing The Power Cord

    Setup 3.5 Installing the Power Cord 1. There are two methods to safely transport the compressor: • One Hand - Grasp the hand grip A at the bottom rear of the compressor. • Two Hands - Grasp the hand grips B on the sides of compressor.
  • Page 16: Connecting The Compressor To The Concentrator

    Invacare® HomeFill® Oxygen System 1. Choose a proper location. Refer to 3.3 Choosing a 1. Insert one end of the interconnect hose C into the Location, page 14. outlet fitting B on the back of the concentrator A. 2. Make sure the power switch is in the Off (O) position.
  • Page 17: Usage

    If filling another cylinder, repeat this checklist. 4.2 Cylinder Prefill Inspection WARNING! – ONLY use cylinders that have the Invacare HomeFill connection and HomeFill label. – All cylinders MUST be inspected before attempting to fill. Otherwise, injury or damage may occur.
  • Page 18: The Hydrostatic Testing Date

    Invacare® HomeFill® Oxygen System The Hydrostatic Testing Date Aluminum cylinders MUST undergo testing every five years. WARNING! – DO NOT fill cylinders that have not been tested in the past five years. Otherwise, serious personal injury may result. Contact your dealer for replacement.
  • Page 19: Connecting The Cylinder To The Compressor

    Usage 4.3 Connecting the Cylinder to the Compressor WARNING! – NEVER use tools of any kind to connect/disconnect the cylinder and the compressor. Otherwise, severe injury and/or damage may occur. WARNING! – DO NOT drop oxygen cylinders. Use two hands when handling/transporting oxygen cylinders.
  • Page 20: Disconnecting The Cylinder From The Compressor

    Invacare® HomeFill® Oxygen System 4.4 Disconnecting the Cylinder from the 3. Remove the cylinder and connector fillport covers (if present). Compressor 4. Momentarily push DOWN on the outer ring (sleeve) B of the connector fillport C until GREEN dots D are visible to reset the connector.
  • Page 21: Turning The Compressor On And Off

    Usage 3. With the other hand, grasp the outer ring (sleeve) of 6. Place the fillport covers onto the connector fillport G the connector fillport E and push DOWN F. and cylinder fillport H. 4. Lift up on the cylinder to remove from the connector When using the post valve type cylinder, fillport.
  • Page 22 Invacare® HomeFill® Oxygen System 1. Make sure the concentrator is On. Refer to the 5. Examine the control panel B. The following sequence concentrator User Manual. of events should occur: a. 0-3 minutes - The O Below Normal (YELLOW) light The maximum oxygen concentrator flow rate C will be lit while the compressor warms up.
  • Page 23: Maintenance

    All other parts and procedures are to be performed by a qualified technician. 5.2 Wear and Tear Invacare reserves the right to ask for any item back that has an alleged defect in workmanship. See the warranty in this manual for specific warranty information.
  • Page 24: Cleaning The Compressor Filter

    Invacare® HomeFill® Oxygen System 5.4 Cleaning the Compressor Filter 3. Clean the filter with a vacuum cleaner C or wash in warm soapy water and rinse thoroughly D. WARNING! 4. Dry the filter thoroughly before reinstallation. – Push the power switch to the Off (O) position and unplug the compressor before cleaning.
  • Page 25: Replacing The Fuse

    Use a 5 mm x 20 mm, time lag, low breaking capacity, 250V, 2 Amp, IEC 60127 style only (Invacare part number 1099820). 6. Push fuse drawer back into the power inlet. 7. Install power cord into the power inlet on the compressor base.
  • Page 26: After Use

    Any accessories not part of the device MUST be handled in accordance with the individual product marking for disposal. Please be environmentally responsible and recycle this product at its end of life. Contact Invacare or your service provider for product recycling information. 1100873-M...
  • Page 27: Troubleshooting

    Troubleshooting 7 Troubleshooting 7.1 Indicator Lights Indicator Light Color Compressor Status Effect None Compressor is Off. flowing to user if concentrator is on. BELOW NORMAL - YELLOW Compressor is On and warming up (0-3 Cylinder filling not started. minutes after power is On). flowing to user if concentrator is On.
  • Page 28 Invacare® HomeFill® Oxygen System Indicator Light Color Compressor Status Effect FULL - GREEN Compressor is On. Cylinder is finished Cylinder filling stops. filling (full) -remove cylinder (FULL light Cannot fill another cylinder until is On in >10 mins.) compressor is turned Off and On again.
  • Page 29: Troubleshooting

    3. Turn the compressor Off. 4. Remove the cylinder and reinstall. 5. Turn the compressor On. If cylinder is not full and is connected properly, the compressor may have internal failure. Contact your home health care provider or Invacare. 1100873-M...
  • Page 30 2. 2. The concentrator may need service, contact your home below normal (yellow) light goes On and Off health care provider or Invacare. below normal (YELLOW) light stays On 1. Ensure the concentrator has warmed up for at least 20 minutes.
  • Page 31 Off, the compressor and/or concentrator may need service. Contact your home health care provider or Invacare. Excessive noise when compressor is On 1. Call home health care provider or Invacare for service. Hissing sound from Regulator when flow selector is Off 1. Turn flow selector Off.
  • Page 32: Technical Data

    VAC/132 VAC), 60 Hz patient therapy flow rate up to 3.0 L/min. Rated Current Input: 2.0 AMP The HomeFill Oxygen System consists of the IOH200 oxygen compressor, connection tubing, and the Invacare supplied Operating Temperature: 50 to 95° F (10 to 35° C) @ portable oxygen cylinders with interlocking fill fitting.
  • Page 33 Technical Data Not suitable for use in the No AP/APG: presence of a flammable anesthetic mixture. Electrical: No extension cords. Placement: No closer than 7.5 cm (3 in) from any wall, furniture, draperies, or furniture to assure sufficient air flow. Avoid deep pile carpets and heaters, radiators or hot air registers.
  • Page 34: Warranty

    Invacare® HomeFill® Oxygen System Invacare or a dealer, with a copy of the seller’s invoice 9 Warranty required for coverage under this warranty, or 5,000 hours, whichever comes first. 9.1 Limited Warranty For Model Nos. IOH2MD, IOH2M6 and IOH2M9, Invacare...
  • Page 35 CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER. shipping damage and to ship the unit to either Invacare's plant or service center as specified by Invacare in advance. SOME STATES DO NOT ALLOW THE EXCLUSION OR...
  • Page 36 Invacare Corporation Manufacturer One Invacare Way Invacare Corporation Elyria, Ohio USA 2101 E. Lake Mary Blvd. 44036-2125 Sanford, FL USA 32773 800-333-6900 407–321–5630 Technical Services: 800-832-4707 www.invacare.com 1100873-M 2018-05-22 Making Life’s Experiences Possible® *1100873M*...