Section 12-Wiring Assemblies; Replacing Wiring Assemblies; Full View Of Wiring Harnesses; Replacing Transformer Assembly - Invacare HomeFillII AW Service Manual

Table of Contents

Advertisement

SECTION 12—WIRING ASSEMBLIES

Replacing Wiring Assemblies

To prevent electrical shock, ALWAYS disconnect compressor from electrical
outlet before servicing.

Full View of Wiring Harnesses

NOTE: For this procedure, refer to FIGURE 12.1.
Bleed Resistor
Assembly 2
Transformer
Assembly Wire
Harness

Replacing Transformer Assembly

NOTE: For this procedure, refer to FIGURE 12.2 on page 82.
Tools required for this procedure are:
• Phillips screwdriver
1. Remove the cabinet. Refer to Removing/Installing the Cabinet on page 25.
2. Remove the control panel. Refer to Removing/Installing Control Panel on page 29.
3. Remove the two mounting screws that secure the transformer to the base (Detail "A").
4. Disconnect the connector from the circuit board (Detail "B").
5. Remove the transformer assembly wire harness from the HomeFill compressor.
Part No 1154316
DANGER
FIGURE 12.1 Full View of Wiring Harnesses
SECTION 12—WIRING ASSEMBLIES
Bleed Resistor
Assembly 1
81
HF Wire
Assembly Harness
NOTE: The bleed
resistors in this view are
for units made before 7/
3/06. On units made
after 7/3/06, bleed
resistors are on the PC
board with a separate
wiring harness.
®
HomeFill
II AW Compressor

Hide quick links: