SECTION 10—COMPRESSOR REPLACEMENT
Timing
Pump Assembly
DETAIL "A"
Large Pulley
Belt
Mounting Screw
Motor Pulley
and Washer
Pump Assembly
Crankshaft
Connecting Lines
Between Stages
DETAIL "C"
DETAIL "B"
Pump
Assembly
Motor
Bushing
Bushing
Pump
Assembly
Mounting
Mounting
Screw
Screw
Washers
FIGURE 10.6 Compressor Replacement - Assembly of New Motor/Pump Assembly

Reassembly

NOTE: For this procedure, refer to FIGURE 10.1 through FIGURE 10.5.
1. Position the motor/pump assembly on the base (FIGURE 10.5).
2. Secure the motor/pump assembly to the base with six mounting screws. Torque
mounting screws to 25-30 in-lbs (2.82-3.39 Nm) (FIGURE 10.5).
3. Position control panel assembly on base (FIGURE 10.4).
4. Connect the two connectors (not shown) to the hour meter (FIGURE 10.4).
®