Configuring for Extended Diagnostics -
XPC/ES8150
When the Scanner is used in the XPC/ES8150 environment, the
Scanner must be connected directly to the Ul Computer in the SIM
before the diagnostics can be used. This subsection describes how to
start the diagnostic software when an ES8150 SIM or a Xerox
Productivity Centre is used as the controlling device for the Scanner.
1. Switch off the System Integration Module (SIM) and the Scanner.
2. (Figure 6-2): Connect the Scanner SCS) Cable directly to the SCSI
port on the User Interface Computer, if not already done during the
installation of the Scanner.
SIM
Figure 6-2. Connecting the Scanner to the SIM
7356 Scanner
Ul
C
Computer
Connect the
Scanner SCSI
Cable to the
SCSI port on
the User
Interface
Computer (Ul)
Disconnect the
SCSI Cable
Disconnect the
Scanner SCSI
Cable
0001075A-COB
3. Ensure that all other SCSI connections are disconnected from the
Scanner.
4. Switch on the Scanner and SIM, and allow the SIM to complete the
boot process.
A
screen
is
displayed
communication with the SIM..." This is normal in diagnostic mode.
5. Press the OK button.
The User Interface screen is displayed.
displayed above the Scanner and Printer icons. This is normal.
6. Enter the Tech Rep mode by pressing <control> <F12>. The
password is 7336884. Press OK after entering the password.
7. Select the Diagnostics button, the OK button, the Scanner Tests
button, the Extended Diagnostics button, and the OK button.
Note: The Normalize Scanner button, displayed above the Extended
Diagnostics button, does not work when the Scanner Is directly
connected to the Ul computer. In this configuration, you must use the
test available inside the extended diagnostics program to normalize the
Scanner.
A screen is displayed, reminding you that "SCSI Connection
from PC to 7356 Required". You have already made this
connection. Press OK.
If the Scanner has not completed its initialization, a screen is displayed
with the message, "The Scanner is warming up. This could take as long
as 3 minutes. If you do not wish to wait for the Scanner to warm up,
select the Exit button to terminate the program."
Once the Scanner has completed its initialization, this screen
disappears.
The software starts.
1. Enter the password ES8150 and press the <Enter> key on the
keyboard.
2. Press the Begin button on the screen.
3. Press the Normalize Cameras button.
At this point you may normalize the Scanner's cameras, or you may
skip the test by pressing Exit. If you plan to perform any diagnostic
tests that require making an image, you should first perform a
camera normalization. Refer to Using the Diagnostic Software for
further information about features of the Diagnostic Software.
4/97
6-3
that
states:
"Could
not
establish
"No SCSI" may be
Diagnostics