1560 Thermometer Readout
User's Guide
5.4.4
56
SELECT A DEVICE TO CALIBRATE
>SPRT 1
TCS
2
PRTS 3
GPIB 4
PRNT 4
AOUT 4
After the device is selected, another window appears showing the parameters
and functions available from the device. These depend on the device. The de-
vice calibration window might appear as follows:
CALIBRATE DEVICE: SPRT
0 ADJ: 0.0
100 ADJ: 0.0
400 ADJ: 0.0
CAL DATE: 05-21-96
SER NUM: 123456
If there are no calibration commands for a particular device the window shows
"NO CALIBRATION OPTIONS FOR THIS DEVICE."
You can scroll through the calibration parameters using the UD buttons. The
values of parameters can be changed using the numeric keys or the LR but-
tons as is appropriate for the parameter type. Press ENTER to set the
parameter.
With function commands you can use the LR buttons to change "NO" to
"YES" and then press ENTER to execute the function. If ENTER is pressed
with "NO" nothing will happen except the cursor will move down to the next
line. Press EXIT to exit.

Module Information

The MODULE INFO soft-key function displays information about a module.
This includes the module name, model number, firmware version number, and
the names of devices it contains with the class and number of channels for each
device (remember a module may contain more than one independent device).
First, you must select the module. A window appears showing a list of mod-
ules. Each module is followed by the module number. The module is selected
using the UD buttons and pressing ENTER.
$