Calibrate Device - Fluke 1560 Black Stack User Manual

Thermometer readout
Table of Contents

Advertisement

SET UP DEVICE: PRNT 4
>LINEFEED: ON
If there are no setup commands for a particular device, the window shows "NO
SETUP OPTIONS FOR THIS DEVICE."
You can scroll through the setup parameters using the UD buttons. The values
of parameters can be changed using the numeric keys or the LR buttons as
is appropriate for the parameter type. Press ENTER to set the parameter.
With function commands you can use the LR buttons to change "NO" to
"YES" and then press ENTER to execute the function. If ENTER is pressed
with "NO" nothing will happen except the cursor will move down to the next
line. Press EXIT to exit.
5.4.3

Calibrate Device

The CAL DEVICE soft-key function is used for accessing module parameters
and functions relating to device calibration. This function may be selectively
locked out using the password feature (see Section 5.5.3 below). If the CAL
DEVICE password option is set ON, in order to access this function you must
enter the correct four-digit password. A screen appears requesting the pass-
word. Use the numeric keys and ENTER to enter the password. If the correct
password is entered, the CAL DEVICE function continues.
A window appears showing a list of devices. The list includes the names of all
devices followed by the position number of the module to which it belongs.
The device is selected using the UD buttons and pressing ENTER.
FORM FEED: NO
PASSWORD REQUIRED FOR ACCESS.
PASSWORD:
5 Soft-Key Functions
LR
Module Menu
55