Operation; Figure 15 Prt Scanner Module Sensor Wiring Detail - Fluke 1560 Black Stack User Manual

Thermometer readout
Table of Contents

Advertisement

1560 Thermometer Readout
User's Guide
9.3
Wiring Detail
4-Wire
1
2
3
4
5
1
Rear Row

Figure 15 PRT Scanner Module Sensor Wiring Detail

126
1
The accuracy specifications apply within the recommended operating temperature range. Accuracy limits
are increased by a factor of the temperature coefficient outside this range.
2
Short-term accuracy includes nonlinearity and noise uncertainties. It does not include drift or calibration
uncertainties.
Add 0.01Ω to accuracy limits when using 3-wire sensors.
3

Operation

Using the 2562/2568 PRT Scanner Module with the 1560 Black Stack is very
simple. The following sections explain the steps for setting up and using the
2562/2568 to measure with PRTs. For instructions on installing the 2562/2568
PRT Scanner Module onto the 1560 see Section 4.1.