Display Data Fields; Data Output Channels - Fluke 1560 Black Stack User Manual

Thermometer readout
Table of Contents

Advertisement

2.3.12

Display Data Fields

In statistical mode, the center display window contains eight programmable
data fields. These can display measurements from any input channel as well as
the results of statistical calculations (see Section 5.3.1.1 and 5.3.1.2). Measure-
ments are displayed with the channel number, units, time, and a label identify-
ing the type of calculation.
2.3.13

Data Output Channels

Output devices can receive measurements from any input channel as well as the
results of statistical calculations (see Section 5.3.2). Each output device pro-
vides one or more output channels. Each device is registered by the base con-
troller on power up and the number of channels are recorded. Output channels
are assigned numbers in the order they are recognized.
2 Introduction
Data Processing
17