Probe Menu; Edit Probe - Fluke 1560 Black Stack User Manual

Thermometer readout
Table of Contents

Advertisement

The AVERAGE parameter determines whether averaging is disabled (OFF) or
enabled (ON). Use the LR buttons to select ON or OFF for AVERAGE and
press ENTER.
The COUNT parameter determines the number of raw measurements that are
averaged to produce the displayed measurement. The range is from 1 to 10. Use
the numeric buttons to enter a value for COUNT and press ENTER.
SET THE INPUT AVERAGING
AVERAGE: OFF
COUNT: 1
5.2

Probe Menu

The PROBE menu provides functions for selecting the type of temperature
conversion and setting the characterization coefficients for input channels. The
soft-key functions that appear in this menu are EDIT PROBE, COPY
PROBE, and TEST CONV. If the PROBE password option is set ON, you
must enter the correct password in order to access these functions (see Section
5.5.3). Use the numeric keys and ENTER to enter the four-digit password. If
the correct password is entered the PROBE menu appears.
5.2.1

Edit Probe

The EDIT PROBE soft-key allows you to enter probe characterization coeffi-
cients and other probe parameters for temperature calculation. First, you must
select the input channel. A window appears showing a list of input channels.
Each channel in the list is identified with the device name, the channel number,
PASSWORD REQUIRED FOR ACCESS.
PASSWORD: 0
5 Soft-Key Functions
Probe Menu
33