Introduction

The Fluke 62 MAX and 62 MAX + Infrared Thermometers (the Product) can determine the surface temperature by measuring the amount of infrared energy radiated by the target's surface. Note that the Japanese models indicate Celsius only.


Read all safety information before you use the Product.

Safety Information

A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test.

Table 1 tells you about symbols used on the Product and in this manual.


To prevent eye damage and personal injury:

  • Read all safety Information before you use the Product.
  • Do not use the Product if it operates incorrectly.
  • Use the Product only as specified, or the protection supplied by the Product can be compromised.
  • Before you use the Product, inspect the case. Do not use the Product if it appears damaged. Look for cracks or missing plastic.

  • See emissivity information for actual temperatures. Reflective objects result in lower than actual temperature measurements. These objects pose a burn hazard.
  • Do not look directly into the laser with optical tools (for example, binoculars, telescopes, microscopes). Optical tools can focus the laser and be dangerous to the eye.
  • Do not look into the laser. Do not point laser directly at persons or animals or indirectly off reflective surfaces.
  • Replace the batteries when the low battery indicator shows to prevent incorrect measurements.
  • Do not use the Product around explosive gas, vapor, or in damp or wet environments.
  • Use the Product only as specified or hazardous laser radiation exposure can occur.

Table 1. Symbols

Symbol Meaning Symbol Meaning
warning Risk of danger. Important information. See Manual. Do not dispose of this product as unsorted municipal waste. Go to Fluke's website for recycling information.
Warning. Laser. Conforms to European Union directives.
Battery Conforms to relevant Australian standards.

Maintenance


To avoid damage to the Product, do not leave the thermometer on or near objects of high temperature.

How to Change the Battery

Installing the battery
To install or change the AA IEC LR06 battery, open the battery compartment and replace the battery as shown in Figure above.

How to Clean the Product

Cleaning the product
Use soap and water on a damp sponge or soft cloth to clean the Product case. Carefully wipe the surface with a moist cotton swab. The swab may be moistened with water.

Specifications

62 MAX 62 MAX +
Temperature Range -30°C to 500°C
(-22°F to 932°F)
-30°C to 650°C
(-22°F to 1202°F)
Accuracy ≥0°C: ±1.5°C or ±1.5% of reading, whichever is greater
(≥32°F: ±3°F or ±1.5% of reading, whichever is greater)
≥ -10°C to <0°C: ±2°C
(≥14°F to <32°F: ±4°F)
< -10°C: ±3°C
(<14°F: ±6°F)
≥0°C: ±1°C or ±1% of reading, whichever is greater (≥32°F: ±2°F or ±1% of reading, whichever is greater)
≥ -10°C to <0°C: ±2°C
(≥14°F to <32°F: ±4°F)
< -10°C: ±3°C
(<14°F: ±6°F)
Response Time (95%) <500 ms (95% of reading) <300 ms (95% of reading)
Spectral Response 8 to 14 microns
Emissivity 0.10 to 1.00
Optical Resolution 10:1 (calculated at 90% energy) 12:1 (calculated at 90% energy)
Display Resolution 0.1°C (0.2°F)
Repeatability (% of reading) ±0.8% of reading or ±1.0°C (2°F), whichever is greater ±0.5% of reading or ±0.5°C (1°F), whichever is greater
Power 1 AA IEC LR06 Battery
Battery Life 10 hours with laser and backlight on 8 hours with laser and backlight on
Weight 255 g (8.99 oz)
Size (175 x 85 x 75) mm (6.88 x 3.34 x 2.95) inches
Operating Temperature 0°C to 50°C (32°F to 122°F)
Storage Temperature -20°C to 60°C (-4°F to 140°F), (without battery)
Operating Humidity 10% to 90% RH non-condensing @ 30°C (86°F)
Operating Altitude 2000 meters above mean sea level
Storage Altitude 12,000 meters above mean sea level
IP Rating IP 54 per IEC 60529
Drop Test 3 meters
Vibration and Shock IEC 68-2-6 2.5 g, 10 to 200 Hz, IEC 68-2-27, 50 g, 11 ms
EMC EN 61326-1:2006 EN 61326-2:2006

Standards and Agency Approval

Compliance - EN/IEC 61010-1: 2001
Laser Safety - FDA and EN 60825-1 Class II

The Product

Using the product - Part 1
Using the product - Part 2
Using the product - Part 3
Using the product - Part 4
Using the product - Part 5
Using the product - Part 6
Using the product - Part 7

LIMITED WARRANTY AND LIMITATION OF LIABILITY

This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. To obtain service during the warranty period, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that Service Center with a description of the problem.

Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.

Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands

How to Contact Fluke

To contact Fluke, call one of the following telephone numbers:

  • Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
  • Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
  • Canada: 1-800-36-FLUKE (1-800-363-5853)
  • Europe: +31 402-675-200
  • Japan: +81-03-6714-3114
  • Singapore: +65-6799-5655
  • Anywhere in the world: +1-425-446-5500

Or, visit Fluke's website at www.fluke.com.
To register your product, visit http://register.fluke.com.

Documents / Resources

References

Download manual

Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.