W(T ) Conversion; Ipts-68 Conversion - Fluke 1560 Black Stack User Manual

Thermometer readout
Table of Contents

Advertisement

5.2.1.3
W(T
90
EDIT SPRT 01
PROBE SER#: 1
CONVERSION: W(T90)
RTPW: 25.412294
The W(T90) conversion displays the measurement in ITS-90 W(T
rather than temperature. The one user-defined parameter for the W(T
sion is the triple point of water resistance, RTPW.
5.2.1.4

IPTS-68 Conversion

EDIT SPRT 01
PROBE SER#: 1
CONVERSION: IPTS-68
SCALE: IPTS-68
R0: 100.00845
ALPHA: 0.00391648
DELTA: 1.4872
The IPTS-68 conversion calculates temperature according to the IPTS-68 speci-
fications. Only subrange 4 is implemented for temperatures below zero. The
user-defined parameters for the IPTS-68 conversion are R0 (R
DELTA (δ), A[4] (A
mines whether the temperature values conform to IPTS-68 or ITS-90.
) Conversion
), C[4], (C
4
$
), and SCALE. The SCALE parameter deter-
4
5 Soft-Key Functions
Probe Menu
) values
90
) conver-
90
), ALPHA α,
0
37