Immunity Testing; Emission Testing; Low Voltage Directive (Safety) - Fluke 1560 Black Stack User Manual

Thermometer readout
Table of Contents

Advertisement

1560 Thermometer Readout
User's Guide
16.6.1.1
16.6.1.2
16.6.2
182
of Conformity for your instrument lists the specific standards to which the unit
was tested.
The instrument was designed specifically as a test and measuring device. Com-
pliance to the EMC directive is through EN 61326-1:2006 Electrical equipment
for measurement, control and laboratory use - EMC requirements
As noted in the EN 61326-1, the instrument can have varying configurations.
The instrument was tested in a typical configuration with shielded RS-232
cables.

Immunity Testing

The instrument was tested to the requirements for industrial locations. This al-
lows the instrument to be used in all types of locations from the laboratory to
the factory floor. Criterion A, B, or C was used for Radiated RF (IEC
61000-4-3) depending on the module configuration. Criterion A or B was used
for Conducted RF (IEC 61000-4-6) depending on the module configuration.
Therefore, the operation of the instrument may be affected by excessive elec-
tromagnetic interference and the instrument may not perform within the normal
specification limits in such an environment.
Criterion A or C was used for Electrostatic Discharge (ESD, IEC 61000-4-2)
depending on the module configuration. Criterion A or B was used for Electric
Fast Transit (EFT, Burst, IEC 61000-4-4) depending on the module configura-
tion. If the instrument is subjected to ESD conditions at industrial levels, the in-
strument may require the user to cycle the power to return to normal operation.

Emission Testing

The instrument fulfills the limit requirements for Class A equipment but does
not fulfill the limit requirements for Class B equipment. The instrument was
not designed to be used in domestic establishments.

Low Voltage Directive (Safety)

In order to comply with the European Low Voltage Directive (73/23/EEC),
Hart Scientific equipment has been designed to meet the IEC 1010-1 (EN
61010-1) and the IEC 1010-2-010 (EN 61010-2-010) standards.

Advertisement

Table of Contents
loading

Table of Contents