Front-Panel Access; Table - Fluke 1560 Black Stack User Manual

Thermometer readout
Table of Contents

Advertisement

Table 29 Themocouple Scanner Module calibration parameters
Parameter
AMP GA
OS 1
GA 1
CJ OS 1
CJ OS 2
CJ OS 3
CJ OS 4
CJ OS 5
CJ OS 6
OS 2
GA 2
CJ OS 7
CJ OS 8
CJ OS 9
CJ OS 10
CJ OS 11
CJ OS 12
CAL DATE
13.4.2

Front-Panel Access

The calibration parameters can be accessed from the front panel of the 1560 us-
ing the CAL DEVICE function in the MODULE soft-key menu (see Section
5.4.3). Select the TCS device. A window appears showing the calibration pa-
rameters for the device. The calibration parameters appear in two screens. The
two screens appear alternately each time the device calibration window ap-
pears. The first screen contains the first nine parameters—AMP GA through CJ
OS 6.
Description
Set to the nominal gain of the amplifier, type 1 or 2
Adjusts the voltage accuracy of Inputs 1–6 at 0 mV
Adjusts the voltage accuracy of Inputs 1–6 at 100 mV
Adjusts the internal CJC accuracy of Input 1
Adjusts the internal CJC accuracy of Input 2
Adjusts the internal CJC accuracy of Input 3
Adjusts the internal CJC accuracy of Input 4
Adjusts the internal CJC accuracy of Input 5
Adjusts the internal CJC accuracy of Input 6
Adjusts the voltage accuracy of Inputs 7–12 at 0 mV
Adjusts the voltage accuracy of Inputs 7–12 at 100 mV
Adjusts the voltage accuracy of Input 7
Adjusts the voltage accuracy of Input 8
Adjusts the voltage accuracy of Input 9
Adjusts the voltage accuracy of Input 10
Adjusts the voltage accuracy of Input 11
Adjusts the voltage accuracy of Input 12
Records the date the module was calibrated
13 2566 Thermocouple Scanner Module
Calibration
159