1560 Thermometer Readout
User's Guide
5.2.1.8
5.2.1.9
40

Thermistor R(T) Conversion

EDIT STHR 01
PROBE SER#: 1
CONVERSION: THRM-R(T)
B[0]: -35.540960
B[1]: 0.36568108
B[2]: -1.884784E-4
B[3]: 7.26691E-6
The following Steinhart-Hart equation is used for the thermistor R(T)
conversion:
=
+
r T K
( [ ])
exp[
B
B T
0
The user-defined parameters for the thermistor R(T) conversion are B[0] (B
through B[3] (B
). If B
3
should be set to 0.0.

Thermocouple Volts Conversion

CHANNEL: TCS 3
PROBE SER#: 1
CONVERSION: V
The volts conversion for thermocouples displays the measurement in volts
rather than temperature.
The voltage readings are not compensated with the CJC value.
Note: When selecting volts as the conversion type for thermocouples, no
cold-junction compensation is available. The value measured and displayed is
the uncompensated voltage as sensed at the input of the module.
+
+
1
2
3
B T
B T
]
1
2
3
is not used, as is the case with some calibrations, B[2]
2
)