Calculate[N]:Average[N]:Clear:all; Calculate[N]:Average[N]:Data - Fluke 1560 Black Stack User Manual

Thermometer readout
Table of Contents

Advertisement

1560 Thermometer Readout
User's Guide
6.5.1.2
6.5.1.3
70
Table 9 Measurement Data Commands
Command
CALCulate[n]:AVERage[n]:CLEar

CALCulate[n]:AVERage[n]:CLEar:ALL

CALCulate[n]:AVERage[n]:DATA?

CALCulate[n]:AVERage[n][:STATe]?
CALCulate[n]:AVERage[n]:TYPE?
CONFigure[<channel>]
CONFigure?
DATA[:DATA]:VALue? [MEM,]<numeric_value>
DATA:POINts?
FETCh[:TEMPerature]?
MEASure[:TEMPerature]? [<channel>]
READ[:TEMPerature]?
SENSe[n]:AVERage:DATA?
given, is ignored as this command clears all the statistical functions for the
specified channel regardless.
Example command: CALC3:AVER:CLE
CALCulate[n]:AVERage[n]:CLEar:ALL
This command simultaneously clears the statistical functions for all channels.
The CALCulate suffix number and AVERage suffix number, if given, are ig-
nored as this command clears all the statistical functions for all channels
regardless.
Example command: CALC:AVER:CLE:ALL
CALCulate[n]:AVERage[n]:DATA?
This query command returns the value of a statistical calculation for a channel.
The CALCulate suffix number specifies the input channel number. Its range is
1 to 96. If it is omitted it is assumed to be channel 1.The AVERage suffix num-
ber specifies the calculation type as shown in Table 10. If it is omitted it is as-
sumed to be type 1 which is the average. The response is a single numeric value
that is the most recent results of the specified calculation for the specified chan-
nel. The following example reads the standard deviation of measurements of in-
put channel 3.
Action
Clear the statistics functions for one channel
Clear the statistics functions for all channels
Return a statistical value for a channel
Return the state of a statistical calculation
Return the type of statistical calculation
Select the input channel to measure
Query measurement function
Return one measurement in memory
Query the memory array size
Return the most recent measurement
Acquire and return one new measurement
Acquire and return one new measurement
Return resistance or voltage

Advertisement

Table of Contents
loading

Table of Contents