System:communicate:serial[:Receive]:Fdup; System:communicate:serial[:Receive]:Linefeed <Boolean; System:communicate:serial[:Receive]:Linefeed; Table 19 Communication Interface Commands - Fluke 1560 Black Stack User Manual

Thermometer readout
Table of Contents

Advertisement

6.5.7.4

SYSTem:COMMunicate:SERial[:RECeive]:FDUP?

This query command returns the duplex or echo mode of the base serial port.
This command returns 1 if full duplex is on and 0 otherwise.

Table 19 Communication Interface Commands

Command
SYSTem:COMMunicate:SERial[:RECeive]:BAUD
<numeric_value>
SYSTem:COMMunicate:SERial[:RECeive]:BAUD?
SYSTem:COMMunicate:SERial[:RECeive]:FDUPlex
SYSTem:COMMunicate:SERial[:RECeive]:FDUP?

SYSTem:COMMunicate:SERial[:RECeive]:LINefeed

<Boolean>
SYSTem:COMMunicate:SERial[:RECeive]:LINefeed?
Example command: SYST:COMM:SER:FDUP?
Example response: 0
6.5.7.5
SYSTem:COMMunicate:SERial[:RECeive]:LINefeed <Boolean>
This command enables or disables linefeed for the base serial port. A
<Boolean> value of 1 or ON turns the linefeed on and 0 or OFF turns it off. If
the linefeed is ON, a linefeed character (ASCII decimal 10) will be appended to
the carriage return at the end of each line during transmission. Linefeed is not
affected by the *RST command.
Example command:SYST: COMM:SER:LIN ON
6.5.7.6
SYSTem:COMMunicate:SERial[:RECeive]:LINefeed?
This query command returns the state of the linefeed for the base serial port. It
returns 1 if the linefeed is on and 0 otherwise.
Example command: SYST:COMM:SER:LIN?
Example response: 1
6 Digital Communications Interface
Action
Set the baud rate for the base serial port
Query the baud rate for the base serial port
Select duplex (echo) mode for the base serial port
Query duplex (echo) mode for the base serial port
Enable or disable linefeed for the base serial port
Query linefeed for the base serial port
Commands
93