2562/2568 Prt Scanner Module; Description; Specifications - Fluke 1560 Black Stack User Manual

Thermometer readout
Table of Contents

Advertisement

9

2562/2568 PRT Scanner Module

This section explains the features and operation of the optional 2562/2568 PRT
Scanner Module.
9.1

Description

The 2562/2568 PRT Scanner Module is an add-on module that allows the 1560
to measure temperature with up to eight SPRT, PRT, and RTD sensors. The
modules accept two-, three-, or four-wire sensors with nominal resistance of 25
and 100Ω (2562) or 500 and 1KΩ (2568). The excitation current alternates po-
larity to minimize the effects of thermoelectric offsets. Up to eight sensors can
be connected to the 2562/2568 simultaneously and measured alternately.
9.2

Specifications

Resistance range
Resistance accuracy, one-year
(Using four-wire connection)
Resistance accuracy, short-term
(Using four-wire connection)
Temperature accuracy, one-year (Using
four-wire connection, not including probe
uncertainty)
–100°C
0°C
100°C
200°C
300°C
400°C
500°C
600°C
Temperature coefficient
Excitation current
Maximum lead resistance
Sample time
Number of channels
Recommended operating temperature
range
1
Absolute operating temperature range
Weight
1, 3
1, 2, 3
1, 3
1
9 2562/2568 PRT Scanner Module
2562
0 to 400Ω
0 to 25Ω: 0.001Ω
25 to 400Ω: 40 ppm of read-
ing
0 to 25Ω: 0.0005Ω
25 to 400Ω: 20 ppm of read-
ing
±0.006°C
±0.010°C
±0.014°C
±0.020°C
±0.024°C
±0.028°C
±0.034°C
±0.040°C
0.5 ppm/°C
1.0, 1.4 mA; 1 Hz
100Ω
2 seconds
18 to 28°C (64 to 82°F)
5 to 35°C (40 to 95°F)
2.5 lbs. (1.1 kg)
Description
2568
0 to 4 kΩ
0 to 250Ω: 0.01Ω
250 to 4KΩ: 40 ppm of read-
ing
0 to 250Ω: 0.0075Ω
250 to 4KΩ: 30 ppm of read-
ing
2.5 ppm/°C
0.1, 0.05 mA; 1 Hz
8
125