System Reset - Fluke 1560 Black Stack User Manual

Thermometer readout
Table of Contents

Advertisement

1560 Thermometer Readout
User's Guide
5.5.5
60
THE
MODEL 1560
HART SCIENTIFIC
FIRMWARE VERSION: 1.20
MODULES:
SPRT-2560
The window shows the product name, model number, manufacturer, and firm-
ware version number. It also shows a list of installed modules. The modules are
listed one per line with the module name and model number. If the list is too
long to fit on one screen the UD buttons can be used to scroll the list. Press
EXIT or ENTER to exit.

System Reset

The SYSTEM RESET function resets operating parameters to default values. It
has the following effects:
• Sets the measure mode to off.
• Sets the COUNT N number to 1.
• Sets the measure delay to 0.
• Sets the primary channel to input channel 1.
• Sets the scan channel list to include all channels.
• Sets the scan mode to primary channel.
• Sets input average to off.
• Disables routing to all output channels by setting the positive input chan-
nel number to 0, the negative input channel number to 0, and the calcula-
tion type to temperature.
• Sets printing to all printer interface devices to off.
• Clears the statistical functions.
• Sets units to C.
Reset does not affect any of the probe characterization parameters, module
setup parameters, or module calibration parameters. A window will appear to
warn you that parameters will be changed. Press ENTER to reset the parame-
ters or press EXIT to cancel.
$