5.2.3

Test Conversion

The TEST CONV soft-key allows you to test the probe characterization algo-
rithm and characterization coefficients for a specific probe. You must first se-
lect the channel number of the probe. A window appears requesting you to
select the input channel. Each channel in the list is identified with the device
name, the channel number, and the probe serial number. The channel is selected
using the UD buttons and pressing ENTER.
SELECT A PROBE TO TEST
>SPRT 01, PROBE 566-011
After the probe is selected, a new window appears allowing you to test the tem-
perature calculation for the probe. You can enter an input value (e.g. resistance)
and the corresponding output value (e.g. temperature) will be shown. Press
EXIT to exit.
TEST CONVERSION
RESISTANCE: 100.0145
TEMP (C): 0.0101