Route:scan:state <Boolean; Route:scan:state - Fluke 1560 Black Stack User Manual

Thermometer readout
Table of Contents

Advertisement

1560 Thermometer Readout
User's Guide
6.5.3.8
6.5.3.9
80
ROUTe:SCAN:STATe <Boolean>
This command enables or disables input channel scanning. It is equivalent to
selecting between the PRIM CHAN and SCAN options with the SCAN
MODE function in the INPUT soft-key menu. A <Boolean> parameter value
of 1 or ON enables scanning. This also disables the alternate scan mode (see
Section 6.5.3.4 ) so that only the selected scan channels are measured. A value
of 0 or OFF disables scanning so that only the primary input channel is mea-
sured. The *RST command sets scanning to OFF.
Example command: ROUT:SCAN:STAT ON

ROUTe:SCAN:STATe?

This query command returns 1 if the scan state is ON and 0 otherwise.
Example command: ROUT:SCAN:STAT?
Example response: 0