Omron V530-R150E-3 Operation Manual page 173

2-dimensional code reader (fixed type)
Hide thumbs Also See for V530-R150E-3:
Table of Contents

Advertisement

SET (Setting) Mode
Note If Display Settings/Finder Pattern is set to ON, whether finder
Procedure
1,2,3...
156
After changing the setting, perform a search to confirm whether
the search can be performed correctly.
Pixel Skipping Y
4.
Set the correlation threshold value for detecting finder patterns.
Correlation values equal to or higher than the threshold are de-
tected. (1 to 100; Default setting: 55.)
5.
Set the timeout time for decode processing of a single image
read. (200 to 2,000 ms; Default setting: 500 ms.)
6.
Select the number of retries to be made if the reading is NG. A
fresh image will be obtained and reading performed for the num-
ber of times specified.
(0 to 9; Default setting: 0.)
patterns can be detected or not can be confirmed on the screen.
“ page 142
• If nothing is detected, decrease the correlation threshold val-
ue and the pixel skipping parameter and check the model im-
age.
• If too many places are detected, increase the correlation val-
ue and pixel skipping parameter.
1.
Select DM (Pattern).
Pixel Skipping X
Set the number of pixels
skipped when searching.
Section 6-1

Advertisement

Table of Contents

Troubleshooting

loading

This manual is also suitable for:

V530-r150ep-3

Table of Contents