Setting Limit Values - Keithley 2016 User Manual

Thd multimeter
Table of Contents

Advertisement

Setting limit values

Use the following steps to enter high and low limit values:
1.
Press the SHIFT-LIMITS keys to view the present HI1 limit value:
HI1:+1.000000
This value represents the absolute value of that function.
2.
Use the
enter the desired value. Move the cursor to the rightmost position (^) and use the
keys to move the decimal point.
3.
Press ENTER to view the present LO1 limit value:
LO1:-1.000000
This value represents the absolute value of that function.
4.
Enter the desired value for this low limit.
5.
Press ENTER to view the present HI2 limits value:
HI2: +2.000000^
This value represents the absolute value of that function.
6.
Enter the desired value for this high limit.
7.
Press ENTER to view the present LO2 limit value:
LO2: -2.000000^
This value represents the absolute value of that function.
8.
Enter the desired value for the low limit. Pressing ENTER returns to the normal display.
^
or