Testing Diodes; Connections; Range; Figure 2-10 Diode Testing - Keithley 2016 User Manual

Thd multimeter
Table of Contents

Advertisement

Testing diodes

With a Model 2016, you can measure the forward voltage drop of general-purpose diodes and
the zener voltage of zener diodes. To test diodes, press SHIFT then
range, connect the diode, and take a reading from the display.
NOTE

Connections

Connect the diode leads to the INPUT HI and INPUT LO terminals on the Model 2016. The
test current flows from the INPUT HI terminal as shown in
Figure 2-10
Diode testing
LOCAL
POWER
LOCAL
POWER
Note: Source current flows from the INPUT HI to INPUT LO terminals.

Range

You can set the test current range from the front panel. The choices are 1mA, 100µA, and
10µA. The factory test current setting is 1mA. To set the test current, do the following:
1.
2.
The diode test measures voltages on the 3V range for the 1mA test current and the 10V range
for the 100µA and 10µA ranges. If a reading is more than 10V, the Model 2016 displays the
"OVERFLOW" status message.
Diode test has a non-selectable reading rate of MEDium (1 PLC).
Model 2016
REM
STEP
SCAN
CH1
CH2
CH3
CH4
CH5
CH6
CH7
CH8
TALK
LSTN
SRQ
SHIFT
TIMER
HOLD
TRIG
FAST
MED
SLOW
REL
FILT
AUTO
ERR
MATH
T H D
dBm
dB
CONT
Ω2
Ω4
SHIFT
DCV
ACV
DCI
ACI
DELAY
HOLD
LIMITS
ON/OFF
TEST
CAL
EX TRIG
TRIG
STORE
RECALL
FILTER
REL
SAVE
SETUP
RS232
CONFIG
HALT
GPIB
SOURCE MEAS
STEP
SCAN
DIGITS RATE
THD
Model 2016
REM
STEP
SCAN
CH1
CH2
CH3
CH4
CH5
CH6
CH7
CH8
TALK
LSTN
SRQ
SHIFT
TIMER
HOLD
TRIG
FAST
MED
SLOW
REL
FILT
AUTO
ERR
MATH
T H D
dBm
dB
CONT
Ω2
Ω4
DCV
ACV
ACI
SHIFT
DCI
DELAY
HOLD
LIMITS
ON/OFF
TEST
CAL
EX TRIG
TRIG
STORE
RECALL
FILTER
REL
SAVE
SETUP
CONFIG
HALT
GPIB
RS232
SOURCE MEAS
STEP
SCAN
DIGITS RATE
THD
Press SHIFT then
.
Use the
and
keys to scroll through the three test current selections.
SENSE
INPUT
Ω 4 WIRE
HI
CH9
CH10
MATH
REAR
4W
350V
1000V
BUFFER
STAT
!
PEAK
PEAK
2016 THD MULTIMETER
LO
500V
PERIOD TCOUPL
PEAK
INPUTS
FREQ
TEMP
RANGE
F
R
AUTO
FRONT/REAR
3A 250V
AMPS
RANGE
EXIT
ENTER
SENSE
INPUT
Ω 4 WIRE
HI
CH9
CH10
MATH
REAR
4W
350V
1000V
BUFFER
STAT
!
PEAK
PEAK
2016 THD MULTIMETER
LO
500V
PERIOD TCOUPL
PEAK
INPUTS
FREQ
TEMP
RANGE
F
R
AUTO
FRONT/REAR
3A 250V
AMPS
RANGE
EXIT
ENTER
Basic Measurements
, set the test current
Figure
2-10.
General-Purpose
Diode
Zener
Diode
2-33

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

20152015p

Table of Contents