Basic Measurements

Measuring current

The Model 2016 can make DCI measurements from 10nA to 3A and ACI measurements from
1µAm to 3A RMS.
NOTE

Connections

Assuming factory default conditions, the basic procedure is as follows:
1.
2.
3.
4.
CAUTION
5.
6.
Figure 2-5
DC and AC current
measurements
See the previous discussion about crest factor in "Measuring voltage" in this section.
Connect test leads to the AMPS and INPUT LO terminals. The front inputs must be used;
place the INPUTS button in the FRONT position.
Select the measurement function by pressing DCI or ACI.
Pressing AUTO toggles autoranging. Notice the AUTO annunciator is displayed with
autoranging. If you want manual ranging, use the RANGE
measurement range consistent with the expected current.
Connect test leads to the source as shown in
Do not apply more than 3A, 250V to the input or the AMPS fuse will
open-circuit.
Observe the display. If the "OVERFLOW" message is displayed, select a higher range
until a normal reading is displayed (or press AUTO for autoranging). Use the lowest
possible range for the best resolution.
Take readings from the display.
Model 2016
REM
STEP
SCAN
CH1
CH2
CH3
CH4
CH5
CH6
TALK
LSTN
SRQ
SHIFT
TIMER
HOLD
TRIG
FAST
MED
SLOW
REL
FILT
AUTO
dBm
dB
MATH
T H D
CONT
Ω2
DCV
ACV
ACI
SHIFT
DCI
HOLD
DELAY
LIMITS
ON/OFF
TEST
LOCAL
EX TRIG
TRIG
STORE
FILTER
RECALL
POWER
SAVE
SETUP
CONFIG
HALT
GPIB
SOURCE MEAS
STEP
SCAN
DIGITS RATE
THD
Caution: Maximum Input = 3A DC or RMS
Figure
SENSE
INPUT
Ω 4 WIRE
HI
CH7
CH8
CH9
CH10
MATH
REAR
4W
350V
ERR
BUFFER
STAT
!
PEAK
2016 THD MULTIMETER
LO
PERIOD TCOUPL
INPUTS
Ω4
FREQ
TEMP
RANGE
F
R
CAL
AUTO
FRONT/REAR
REL
3A 250V
AMPS
RANGE
RS232
EXIT
ENTER
and
keys to select a
2-5.
1000V
PEAK
Current
500V
Source
PEAK

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

20152015p

Table of Contents