:OUTPut:LIST <ampl#1-NRf, freq#1-NRf,ampl#2-NRf, freq#2-NRf... >
Parameter
<ampl#1-NRf>
<freq#1-NRf>
Query
:LIST?
Description
Sets the sweep list points. This command restarts the list from point #1 when
sent. If you want to sweep more than 50 points you must use the
OUTP:LIST:APPend command to set points 51 and above up to a maximum
of 200 total points.
:OUTPut:LIST:APPend <ampl#X-NRf, freq#X-NRf... >
Parameter
<freq#X-NRf>
Description
Append points to the end of the sweep list (maximum of 50 points at a time)
up to a maximum of 200 points total.
:OUTPut:LIST:DELay <NRf>
Parameter
<NRf>
Query
:DELay?
Description
Set the source settling delay in seconds.
:OUTPut:LIST:ELEMents <DISTortion, AMPLitude>
Parameters
<DIST>
<AMPL>
Query
: ELEMents?
Description
Set reading elements to be returned by OUTP:LIST:DATA?. Any
combination of the above can be used. This command works like
FORM:ELEM.
:OUTPut:LIST:DATA?
Description
Query the results of the sweep (in the form DIST1, DIST2, etc., or AMPL1,
AMPL2, etc., or DIST1, AMPL1, DIST2, AMPL2, etc., depending on
ELEMents previously set. This command will return zero's if the ELEMents
have been changed since the last sweep was performed (this is to ensure that
the dist readings and ampl readings are synchronized pairs).
NOTE
:OUTPut:LIST:DATA? is a non-overlapped command. Once initiated, the sweep will
continue until the sweep done bit is asserted (swd). For more information on the
sweep done bit, refer to the operation event status contained in Section 4.
SCPI Command Reference
Amplitude list point (50 maximum).
Frequency list point (50 maximum).
Query sweep points.
Frequency list point (50 maximum per APPend).
0 (default) to 999999.999s
Query source settling time.
Distortion value.
Amplitude list point.
Query the reading elements to return by OUTP:LIST:DATA?.
5-35