Elements Command - Keithley 2016 User Manual

Thd multimeter
Table of Contents

Advertisement

:ELEMents command

:ELEMents <item list>
:FORMat:ELEMents <item list>
Parameters
Query
Description
<item list>: READing
CHANnel
UNITs
NOTE: Each item in the list must be separated by a comma (,).
:ELEMents?
Query elements in data string
This command is used to specify the elements to be included in the data
string for each measurement conversion. You can specify from one to all
three elements. Each element in the list must be separated by a comma (,).
These elements, shown in
READing: Instrument reading. The resolution of this reading tracks the
display resolution of the instrument. An overflow reading reads as +9.9e37
with no units.
CHANnel: Corresponds the instrument reading to the channel number of a
switching card. If not scanning, the channel number is 0.
UNITs: This element attaches the function unit to the reading, and the
channel unit (internal or external) to the channel number. An internal channel
refers to an internally installed switching card channel, while an external
channel refers to the channel for an external switch system. This element is
not available for the binary formats.
The ASCII format shown in
string. Keep in mind that the byte order can only be reversed for the binary
formats.
When using this command to add an element, you must include all elements
that you want in the format. For example, if the reading is already specified
and you want to add the channel, you must include the READing parameter:
:form:elem chan, read
Data elements for the item list can be listed in any order, but are always sent
in the order shown in
Figure
SCPI Command Reference
Includes reading in data string
Includes channel number
Includes units
Figure
5-1, are explained as follows:
Figure 5-1
shows the byte order of the data
5-1.
5-33

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

20152015p

Table of Contents