5-40
SCPI Command Reference
:APERture <n>
[:SENSe[1]]:FREQuency:APERture <n>
[:SENSe[1]]:PERiod:APERture <n>
Parameters
Query
Description

:RANGe commands

[:UPPer] <n>
[:SENSe[1]]:CURRent:AC:RANGe[:UPPer] <n>
[:SENSe[1]]:CURRent[:DC]:RANGe[:UPPer] <n>
[:SENSe[1]]:VOLTage:AC:RANGe[:UPPer] <n>
[:SENSe[1]]:VOLTage[:DC]:RANGe[:UPPer] <n>
[:SENSe[1]]:RESistance:RANGe[:UPPer] <n>
[:SENSe[1]]:FRESistance:RANGe[:UPPer] <n>
[:SENSe[1]]:DISTortion:RANGe[:UPPer] <n>
Parameters
Query
Description
<n> = 0.01s to 1.0s
DEFault
1.0s
MINimum
0.01s
MAXimum 1.0s
:APERture?
Query programmed APERture value
The integration period (measurement speed) for Frequency and Period is set
using the APERture command (to set other basic measurement functions, see
:NPLCycle). The unit of measurement for the APERture command is
seconds.
<n> =
0 to 3.1
0 to 757.5
0 to 1010
0 to 120e6
0 to 757.5
DEFault
MINimum
MAXimum
:RANGe[:UPPer]?
:RANGe[:UPPer]? DEFault
:RANGe[:UPPer]? MINimum
:RANGe[:UPPer]? MAXimum
This command is used to manually select the measurement range for the
specified measurement function. The range is selected by specifying the
expected reading as an absolute value. The Model 2016 will then go to the
most sensitive range that will accommodate that expected reading. For exam-
ple, if you expect a reading of approximately 50mV, simply let the parameter
(<n>) = 0.05 (or 50e-3) in order to select the 100mV range.
Set aperture for frequency
Set aperture for period
Set measurement range for ACI
Set measurement range for DCI
Set measurement range for ACV
Set measurement range for DCV
Set measurement range for Ω2
Set measurement range for Ω4
Set measurement range for distortion
Expected reading is amps (ACI and DCI)
Expected reading is AC volts (ACV)
Expected reading in DC volts (DCV)
Expected reading is ohms (
Expected input signal is AC volts (distortion)
3 (ACI and DCI)
757.5 (ACV)
1000 (DCV)
100e6 (
2 and
4)
0 (All functions)
Same as DEFault
Query measurement range
Query *RST default range
Query lowest measurement range
Query highest measurement range
2 and
4)