:STATe <b>
[:SENSe[1]]:HOLD:STATe <b>
Parameters
Query
Description

Speed commands

:NPLCycles <n>
[:SENSe[1]]:CURRent:AC:NPLCycles <n>
[:SENSe[1]]:CURRen[:DC]:NPLCycles <n>
[:SENSe[1]]:VOLTage:AC:NPLCycles <n>
[:SENSe[1]]:VOLTage[:DC]:NPLCycles <n>
[:SENSe[1]]:RESistance:NPLCycles <n>
[:SENSe[1]]:FRESistance:NPLCycles <n>
[:SENSe[1]]:TEMPerature:NPLCycles <n>
Parameters
Query
Description
NOTE
:SENS:VOLT:AC:NPLC and :SENS:CURR:AC:NPLC are valid only for a bandwidth
of 300Hz.
<b> =
0 or OFF
1 or ON
:STATe?
Query state of Hold.
This command is used to enable or disable Hold. See "Hold" in Section 3 and
"Trigger Model, Device Action" in this section for details on Hold.
<n> =
0.01 to 10
DEFault
MINimum
MAXimum 10
:NPLCycles?
:NPLCycles? DEFault
:NPLCycles? MINimum
:NPLCycles? MAXimum
The integration period (measurement speed) for the basic measurement
functions (except Frequency and Period) is set using the :NPLCycle
command. NPLC (Number of Power Line Cycles) expresses the integration
period by basing it on the power line frequency. For example, for a PLC of
1, the integration period in seconds would be 1/60 (for 60Hz line power)
which is 16.67msec.
SCPI Command Reference
Control (on/off) Hold
Disable Hold
Enable Hold
Set NPLC for ACI
Set NPLC for DCI
Set NPLC for ACV
Set NPLC for DCV
Set NPLC for Ω2
Set NPLC for Ω4
Set NPLC for TEMP
Power line cycles per integration
1
0.01
Query programmed NPLC value
Query *RST default value
Query minimum NPLC value
Query maximum NPLC value