:DB:REFerence <n>
:UNIT:VOLTage:AC:DB:REFerence <n>
Parameter
<n> = 1e-7 to 1000
Query
:REFerence?
Description
This command is used to specify the dB reference level. When DB units is
selected (:VOLTage:AC: DB), ACV dB measurements are made using the
specified dB reference level.
The reference level is specified in volts and is not range dependent. For
example, a dB reference level of 1 is 1V on all ACV measurement ranges.
:DBM:IMPedance <n>
:UNIT:VOLTage:AC:DBM:IMPedance <n>
Parameter
<n> = 1 to 9999 Specify reference impedance
Query
:IMPedance?
Description
This command is used to specify the dBm reference impedance level. When
dBm units is selected, ACV dBm measurements are made using the specified
dBm reference impedance.
The reference impedance is specified in ohms and is not range dependent.
For example, a dBm reference level of 600 is 600 on all ACV measurement
ranges. A rational number is rounded to the nearest valid integer value.
[:DC] <name>
:UNIT:VOLTage:DC <name>
Parameters
<name> =
Query
[:DC]?
Description
This command is used to select the units for DCV measurements. With volt
(V) units selected, normal DC voltage measurements are made for the DCV
function. With DB units selected, DC dB voltage measurements are
performed. The DBM units selection is used to make decibel measurements
referenced to 1mW. dB and dBm measurements are explained further in
Section 2.
SCPI Command Reference
Specify dBm reference
Specify reference in volts
Specify dB reference
Specify DCV units
V
DC voltage measurement units
DB
dB DC voltage measurement units
DBM
dBm DC voltage measurement units
Query DC voltage units