2105 Universal Cable 60-2004-580; Preparation; 2105 Connection - Teledyne 2105 Installation And Operation Manual

Interface module
Table of Contents

Advertisement

2105 Interface Module
Section 2 Installation and Operation
2.16 2105 Universal Cable
60-2004-580

2.16.1 Preparation

2.16.2 2105 Connection

Figure 2-39 2105 Interface connector
2-38
This multipurpose cable connects to the 12-pin female connector
on the side of the 2105 module. The non-terminated end can be
wired to compatible circuits for interfacing. These instructions
explain how to prepare the cable for use, and the function of each
wire.
Risk of equipment damage. Only experienced electronic tech-
nicians should make the connections to an external device
using cable 69-2004-580.
1. Referring to Table 2-4, select the appropriate wires and tin
the ends with solder or attach crimped connector ends.
2. Electrically insulate any unused wires.
3. Ensure that the connection has protection from the envi-
ronment, such as cable conduit. Teledyne Isco offers a
sealed, in-line cable joiner for water protection.
Ensure that no wire ends come into contact with each other.
Personal injury, power shutdown, or component failure can
result. Wires not in use should be individually sealed at all
times.
To connect the cable to a 2105 module, simply press the cable
connector into the module's connector port until it clicks (Figure
2-39). To disconnect the cable, press down on the spring-loaded
connector tab and pull the cable connector out.
Note
Always insert connector caps into unused connectors to termi-
nate the network and prevent moisture damage.
(Modem
Module Shown)
Universal Cable
Connector
CAUTION
WARNING
Module Connector
Tab