2.12 Sampler "Pass Thru"
Mode
To record sample events and bottle numbers, select Set Up Data
Storage... on the Sampler tab. In the Data Storage Setup window
(Figure 2-28), select the check box next to Enable Logging.
The 2105 module offers a "Pass Thru" mode which, when acti-
vated, allows you to remotely access a 6712 or Avalanche
sampler's configuration settings.
The "Pass Thru" feature allows full remote control of a 6712's (or
Avalanche's) keypad functions. With this access, you can view or
modify all program and configuration settings, and also view
stored reports, current status, and data values. You can start a
program, modify the current program, or stop an active program.
This feature is helpful for applications such as storm monitoring
where the need to adjust program operation due to pending
events is often necessary.
Note
If you use this feature, it is important that you follow the steps
in Table 2-2. Make sure you close all active menus and close
the terminal emulation connection to the sampler, or the 2105
will remain in Pass Thru mode. While in Pass Thru mode, the
2105 will be unable to push data, send alarms, or accept
incoming communications.
If you are using CSD service and need to terminate the Pass
Thru mode, log back into the terminal emulation program and
follow steps 10-14 in Table 2-2.
If you are using SOIP service and you lose communications
with the module while in Pass Thru mode, you will need to visit
the site and cycle the power to restore wireless communica-
tions.