Page 2
Clarity ® , DataApex ® and ® are trademarks of DataApex Ltd. Microsoft ® and Windows TM are trademarks of Microsoft Corporation. DataApex reserves the right to make changes to manuals without prior notice. Updated manuals can be downloaded from www.dataapex.com. Author: DM...
Contents 1 Teledyne Foxy 200 Control Module 2 Requirements 3 Installation Procedure 3.1 Teledyne Foxy 200 fraction collector communication 3.2 Clarity Configuration 4 Using the control module 4.1 Method Setup - FC - Fraction Table 4.2 Method Setup - FC - Vial Numbers 4.3 Hardware Configuration...
Page 4
Table of Contents Clarity Controls To facilitate the orientation in the Teledyne Foxy 200 manual and Clarity chromatography station, different fonts are used throughout the manual. Meanings of these fonts are: Instrument (blue text) marks the name of the window to which the text refers.
Teledyne Foxy 200 1 Teledyne Foxy 200 Control Module 1 Teledyne Foxy 200 Control Module This manual describes the setting of the Teledyne Foxy 200 fraction collector. The control module enables direct control of the instrument over serial line. Fig 1: Teledyne Foxy 200 fraction collector Direct control means that the fraction collector can be completely controlled from the Clarity environment.
Teledyne Foxy 200 2 Requirements 2 Requirements Free serial COM port in the PC. Note: Modern computers usually have only one (if any) serial (COM) port installed. To use more devices requiring the RS232 port, the MultiCOM adapter (p/n MC01) is available.
3 Installation Procedure 3 Installation Procedure 3.1 Teledyne Foxy 200 fraction collector communication The Teledyne Foxy 200 is controlled by serial (RS232) communication. It uses a serial DB9F-DB25M straight cable (p/n SK12) described in the picture below. Fig 2: Serial DB9F-DB25M cable On top of that, the fraction collector must be switched to the same Baud Rate as set in Clarity and the ID=5 to work with the control module.
3 Installation Procedure Clarity Controls 3.2 Clarity Configuration Fig 3: How to Add UNI Ruby module Start the Clarity station by clicking on the icon on the desktop. Invoke the System Configuration dialog accessible from the Clarity window using the System - Configuration... command. Press the Add button (①...
Page 9
Note: DataApex UNI Setup dialog is described in detail in the chapter "DataApex UNI Setup" on pg 18. ④ The Teledyne Foxy 200 item will appear in the Setup Control Modules list of the System Configuration dialog. Drag the appropriate item from the Setup Control Modules list on the left ⑤...
After adding and setting up the new device one or more new tabs will appear in the Method Setup dialog depending on the type of the instrument. A new Teledyne Foxy 200 fraction collector section enabling the monitoring of the current fraction collector state will be also created in Device Monitor window.
4.1 Method Setup - FC - Fraction Table Method Setup - FC - Fraction Table tab is used for setting the automatic fraction collection program of the Teledyne Foxy 200 fraction collector. The actions of the fraction collector can be displayed in Data...
Page 12
Teledyne Foxy 200 4 Using the control module tubing volume between the detector and fraction collector tray should be typed into the Delay Volume Detector - Valve field. Fraction Table Allows to set the fraction collector behavior. The table rows are run in time order, the time intervals should not overlap.
Page 13
4 Using the control module Clarity Controls starts when the detector signal increases above the Level Start stops when the detector signal decreases below the Level Stop Slope - The fraction collection: starts when the first derivative of the detector signal increases above the Slope Start stops when the first derivative of the detector signal first decreases and then increases above the negative of the Slope Stop.
Page 14
Teledyne Foxy 200 4 Using the control module above the negative of the Slope Stop value. If only one of the stop signal conditions is fulfilled the fraction collection will not stop. Start L+S End L or S - The fraction collection:...
Page 15
4 Using the control module Clarity Controls performed, value of X means each X data points received will be averaged to 1 point used for fraction start/end conditions detection. The reaction to the fulfilled start/stop signal condition will be delayed the higher the Filter value is set, mainly on detectors with low data rate.
Open Instrument - the vial number is reset only once, when the Instrument is opened. Last Vial Number Vial Number Sets the number of the last available vial. The Teledyne Foxy 200 fraction collector can set the tray used in the DataApex UNI Setup dialog, so the value range of this field is limited.
Page 17
4 Using the control module Clarity Controls After Last Vial Specifies the action executed when current vial number exceeds the Last Vial Number. Waste - the acquisition continues, but the fraction table is ignored and the samples are automatically discarded. Shutdown - the acquisition is immediately stopped and the analysis aborted.
Hardware Configuration dialog (invoked by using the FC Status button from the Method Setup - FC dialog) displays the configuration of the Teledyne Foxy 200 , namely the communication type and its parameters. Fig 8: Hardware Configuration - 14 -...
4 Using the control module Clarity Controls 4.4 Device Monitor The window with the fraction collector status can be invoked by the Monitor - Device Monitor command from the Instrument window or using Device Monitor icon. Device Monitor - Fraction Collector Fig 9: Device Monitor - Fraction Collector Current Position Displays whether the fraction collector is currently wasting or the position...
Page 20
Teledyne Foxy 200 4 Using the control module moment, it will start a waste cycle and then resume the function of the fraction table. - 16 -...
4 Using the control module Clarity Controls 4.5 Data Acquisition window Fig 10: Data Acquisition window Data Acquisition window can display the fractions in the graph using the background color and also the start and stop fraction events and marker of the Fraction Delay. To see the fractions and markers right click on the graph, select Properties…...
Displays the subtype of the fraction collector. Baud Rate Sets the Baud Rate communication parameter for the Teledyne Foxy 200 fraction collector. Available values are 110, 300, 600, 1200, 2400, 4800, 9600 , 19200 and 38400 baud. Please note that the value set in the...
Page 23
Rack Selection Allows the user to select the correct rack type installed on the Teledyne Foxy 200 fraction collector. The Teledyne Foxy 200 fraction collector supports two racks but both settings can be read from the fraction collector itself by using the Autodetect button.
Teledyne Foxy 200 5 Report Setup 5 Report Setup The fraction collector section on the method report can be enabled by checking the Instrument Control checkbox on the Method tab of the Report Setup dialog. Fig 12: Report Setup All of the parameters set in the...
Note: Instead of COM1, type the communication port used to communicate with the Teledyne Foxy 200 fraction collector. This port number is displayed when the FC Status button in the Method Setup - FC dialog is invoked. Note: %D (or %d) in the filename parameter means that the log will be created separately for each day.