Retrofi Tting Components; Fitting The Filter Magazine (Transmitted Light/Incident Light); Equipping The Condenser Disk - Leica DM2700 M Instructions Manual

Hide thumbs Also See for DM2700 M:
Table of Contents

Advertisement

Available languages

Available languages

14. Retrofi tting Components
14. Retrofitting Components
14.1 Fitting the Filter Magazine
(Transmitted Light/Incident Light)
• Remove the tube and intermediate systems
where applicable.
• Turn the microscope stand/incident light axis
upside down, loosen the fastening screws at
the bottom and lift off the base plate.
• Insert the fi lters into the semicircular mounts.
This does not have to be done in any particular
order.
• Put the fi lter magazine back in position.
Fig. 68
Transmitted light fi lter magazine
for Leica DM2700 M
72

14.2 Equipping the Condenser Disk

• Turn the stage upward and lower the condens-
er.
• Remove the condenser. Therefore loosen the
condenser's clamping screw.
Condenser UCL/UCLP
• Remove the screw (69.1) completely.
• Turn back the centering screws until the light
rings, λ- and λ/4 - compensator* and lens* 2.5x
can be inserted.
The largest hole is for brightfi eld observation
(= BF), the slightly smaller ones for light rings
or λ- and λ/4 - compensator or lens* 2.5x.
Fig. 69
Condenser UCL
1
Fixing screw for condenser disk
1

Hide quick links: