5. Unpacking the Microscope

First, carefully remove all components from the
transportation and packaging materials.
Note:
If at all possible, avoid touching the lens surfaces
of the objectives. If fi ngerprints do appear on the
glass surfaces, remove them with a soft leather
or linen cloth. Even small traces of fi nger perspi-
ration can damage the surfaces in a short time.
See the chapter on „Care of the Microscope" →
p. 69 for additional instructions.
Caution!
Do not connect the microscope and periph-
erals to the power supply at this point under
any circumstances!
Installation location
Work with the microscope should be performed
in a dust-free room, which is free of oil vapors
and other chemical vapors, as well as extreme
humidity. At the workstation, large temperature
fl uctuations, direct sunlight and vibrations should
be avoided. These conditions can distort meas-
urements and micrographic images.
5. Unpacking the Microscope
Allowable ambient conditions:
Temperature
Relative humidity:
Microscopes in warm and warm-damp climatic
zones require special care in order to prevent
fungus contamination.
See the chapter on „Care of the Microscope" →
p. 69 for additional instructions.
Caution!
Electrical components must be assembled
at least 10 cm from the wall and away from
fl ammable substances.
Caution!
When installing the microscope, make sure
the power inlet is freely accessible so that
the instrument can be quickly disconnected
from the mains if necessary.
15–35°C
max. 80% up to 30°C
non-condensing
19