6.8 Analyzer and Polarizer
Analyzer
• Remove the plug cap on the left side of the
stand.
• Insert the analyzer into the receptacle until it
latches in place (24.1).
When using the Pol intermediate tube* or the TL
analyzer slot*:
• Remove the plug cap on the left side.
• Insert the analyzer into the receptacle until it
latches into place.
Fig. 23
Assembling the polarizer holder
1
Clamping screw
Fig. 24
Assembling the analyzer
1
Analyzer
Polarizer
• Attach the polarizer holder to the underside
of the condenser holder with the left clamp
screw (23.1). Remove the fl ip-out blue fi lter if
required.
• Push the polarizer with the labeled side facing
upward into the lower opening.
6.9
Lambda Plate Compensator*
• Raise the condenser to it's upper stop position.
• Remove the DLF fi lter magazine from the base
if present.
• Attach the lambda plate compensator to the
base.
6.10 ICR/ICT Prisms
The condenser prisms have already been mount-
ed at the factory.
To adjust the condenser prisms during fi rst use
→ p. 39.
To retrofi t ICR/ICT prisms, see → p. 72.
2
1
6. Assembling
31