Trouble Shooting - Leica DM2700 M Instructions Manual

Hide thumbs Also See for DM2700 M:
Table of Contents

Advertisement

Available languages

Available languages

11. Trouble Shooting

Problem
Stand
The microscope does not respond.
Illumination
The image is completely dark.
The image is unevenly or not uniformly illumi-
nated.
Cause/Remedy
Make sure that voltage is available.
Make sure that the stand is connected to the
power supply.
Check the cable connections.
Check whether the fuse is defective and re-
place it if necessary (→ p. 70).
Ensure that the lamp/LED in the lamp housing
is not defective and that the lamp housing is
connected to the power supply.
Lamp replacement for LH 107/2 → p. 26f
Lamp replacement for LH 106z → p. 27f
For defective LED, replace lamp housing
Incident Light/Fluorescence
Swing in the brightfi eld cube.
For supply units: Inform Leica Service and have
the supply unit fuse checked.
Remove all unneeded fi lters from the light path.
If present, center the lamp (lamp housing 106z)
(→ p. 41ff).
If present, replace the old lamp (lamp housing
107/2 or 106z) (→ p. 26ff).
11. Trouble Shooting