6. Assembling

6.7 Equipping the Incident Light Turret Disk

Insert the fi lter and refl ector cubes in the follow-
ing manner:
• Remove the analyzer (20.1).
• Remove the front cover (20.2) by pulling it to-
ward the front.
• Insert the fi lter or refl ector cube into the
mounting in front of you.
To do so, place the fi lter or refl ector cube on
the right side and press it to the left or screw it
into the mounting.
Note:
The numbering is located directly below the
holder.
• Affi x the provided stickers (20.3) to the front
side of the incident light axis according to the
equipment.
• When all fi lters and refl ector cubes have been
inserted, close the front cover plate again. En-
sure that the cover snaps into place.
Fig. 18
Filter cube,
front side
30
Fig. 19
Filter cube,
back side
Note:
For the 4x fi lter turret disc, two fi lter cubes for
brightfi eld and darkfi eld have already been
screwed in at the factory. Therefore, two posi-
tions are available to be equipped as desired.
Fig. 20
Incident light and fl uorescence axis
1
Analyzer
2
Front panel
3
Sticker
1
2
Fig. 21
Incident light/fl uorescence axis without front
panel
Fig. 22
Inserting the fi lter cube