6.13 Equipping the Incident Light Turret Disk

!
Caution:
Please read this section completely before be-
ginning with the assembly of the turret disk.
Leica DMI4000 B and Leica DMI6000 B:
The fl uorescence drawer is located on the right
side of the stand. Before opening this drawer,
remove the cap below the drawer covering the
analyzer slot (65.1). Remove the analyzer if it is
already in the slot.
The replacement of individual cubes is more con-
venient with the microscope switched on. The
position to be changed then automatically turns
to the outside and you can be sure that the cube
is positioned in the correct holder. You can there-
fore postpone installing the fi lter cubes until after
the microscope has been switched on.
You can also insert the fi lter cubes while the in-
strument is switched off.
Press the white button next to the drawer. The
drawer will glide out into its initial position.
Fig. 64a
Filter cube,
front side
Fig. 64b
Filter cube,
back side
Fig. 65
Opening the fl uorescence drawer
1 Analyzer slot
Fig. 66
Open fl uorescence drawer
1 Lever for fi xing the loading position
Fig. 67
Inserting or removing a fi lter cube
6. Assembly
1
1
53